User manual

Table Of Contents
DO NOT change the values of the calibration constants from the factory
set values. The correct setting of these parameters is important to the
safety and proper operation of the calibrator.
DO use a ground fault interrupt device.
Operate the instrument in room temperatures between 5–50°C
(41–122° F). Allow sufficient air circulation by leaving at least 6 inches
of space between the instrument and nearby objects.
The instrument is a precision instrument. Although is has been de
-
signed for optimum durability and trouble free operation, it must be
handled with care. The convenient fold-up handle allows one hand
carrying. The instrument should not be operated in excessively wet,
oily, dusty, or dirty environments. It is important to keep the calibra
-
tion probe well and the IR target surface clean and clear of any foreign
matter. Do not operate near flammable materials.
DO NOT use fluids to clean out the calibration probe well or on the IR
target surface.
Before initial use, after transport, and anytime the instrument has not
been energized for a “dry-out” period of 1-2 hours before it can be as-
sumed to meet all of the safety requirements of the IEC 1010-1.
If a main supply power fluctuation occurs, immediately turn off the in-
strument. Wait until the power has stabilized before re-energizing the
instrument.
1.3 Fluke Hart Scientific Authorized
Service Centers
Please contact one of the following authorized Service Centers to coordinate
service on your Hart product:
Fluke Hart Scientific
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
Fluke Nederland B.V.
Customer Support Services
Science Park Eindhoven 5108
5692 EC Son
9133
User’s Guide
4