User manual

3
1 Before You Start
Symbol Description
Canadian Standards Association
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 refers to
the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE
CATEGORY II is energy-consuming equipment to be supplied from the fixed installation.
Examples include household, office, and laboratory appliances.
C-TIC Australian EMC mark
The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/
EC) mark.
1.3 Safety Information
Use this instrument only as specified in this manual. Otherwise, the protection provided by the
instrument may be impaired.
The following definitions apply to the terms “Warning” and “Caution”.
“WARNING” identifies conditions and actions that may pose hazards to the user.
“CAUTION” identifies conditions and actions that may damage the instrument being
used.
1.3.1 WARNINGS
To avoid personal injury, follow these guidelines.
BURN HAZARD – DO NOT touch the IR target surface of the unit.The temperature
of the IR target surface is the same as the actual temperature shown on the display. If
the unit is set at 500°C and the display reads 500°C, the target surface is 500°C.The top
sheet metal of the instrument may exhibit extreme temperatures for areas close to the IR
target surface. DO NOT turn off the unit at temperatures higher than 100°C. This could
create a hazardous situation. Select a set-point less than 100°C and allow the unit to cool
before turning it off.
DO NOT operate this unit without a properly grounded, properly polarized power cord.
DO NOT connect this unit to a non-grounded, non-polarized outlet.
HIGH VOLTAGE is used in the operation of this equipment. SEVERE INJURY OR
DEATH may result if personnel fail to observe safety precautions. Before working inside
the equipment, turn the power off and disconnect the power cord.
Always replace the fuse with one of the same rating, voltage, and type.
This instrument is intended for indoor use only.
Overhead clearance is required. DO NOT place this instrument under a cabinet or other
structure.