Calibration Procedure
12.2 Calibration Procedure 12
12-3
where
T
1
and R
1
are the measured temperature and resistance at 50.0 °C
T
2
and R
2
are the measured temperature and resistance at 90.0 °C
T
3
and R
3
and R are the measured temperature and resistance at 150.0 °C
12.2.2 Compute R0 & ALPHA
+=
100
1
100
11
11
TT
deltaTa
+=
100
1
100
33
33
TT
deltaTa
31
3113
aa
aRaR
rzero
=
=
3113
31
aRaR
RR
alpha
=
=
Where:
delta is the new value of DELTA computed above
Program the new values for DELTA (delta), R0 (rzero), and ALPHA (alpha) into
the Micro-Bath with the following steps.
Press “SET” and “EXIT” keys at the same time and then press “UP” until
“CAL” is displayed.
Press “SET” five times until “R0” is displayed.
Press “SET” then use the “UP” or “DOWN” keys until the correct numerical
setting is displayed. Press “SET” to accept the new value.
Repeat step 2 for ALPHA and DELTA.
12.2.3 Accuracy & Repeatability
Check the accuracy of the Micro-Bath at various points over the calibrated range.
If Micro-Bath does not pass specification at all set-points, repeat the Calibration
Procedure.