Specications and Environmental Conditions
Specications
Specications and Environmental Conditions3
3.1 Specications
The 6102 may not meet the specification published if a different fluid is used other
than distilled water or Dow Corning (200.10, 200.20) oil.
Do not exceed the boiling point or flash point temperatures of the fluid. To ensure that
the boiling and flash point temperatures are not exceeded adjust the High Limit and
Cutout parameters. To set these parameters, refer to Section 9, Controller Operation.
Specifications Table 2
Range 35 to 200°C (95 to 392°F)
Accuracy ±0.25°C
Stability ±0.02°C at 100°C
±0.03°C at 200°C
Uniformity ±0.02°C
Resolution 0.01°C/F
Operating Temperature 5 to 45°C (41 to 113°F)
Heating Time 25°C to 200°C: 40 minutes
Cooling Time 200°C to 100°C: 35 minutes
Well Size 1.9" dia. x 5.5" deep (48 x 139 mm)
Size 5.5" W x 10.38" H x 8" D (14 x 26 x 20 cm)
Weight 10 lb. (4.5 kg) with uid
Power 115 VAC (±10%), 12.3 A, or 230 VAC (±10%), 1.1 A, switchable, 50/60 Hz,
270 W
System Fuse 115 V: 250 V, 3A, F
230 V: 250 V, 1.6A, F
Safety Overvoltage (Installation) Category II, Pollution Degree 2 per IEC 61010-01
3.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trou ble-free
operation, it must be handled with care. The instrument should not be operated in an
excessively dusty or dirty environment. Maintenance and clean ing recommendations
can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
ambient temperature range: 5 – 45°C (41 – 113°F)
L
ambient relative humidity: maximum 80% for temperature <31°C, de creasing
L
linearly to 50% at 40°C
mains voltage within ± 10% of nominal
L
vibrations in the calibration environment should be minimized
L
altitudes less than 2,000 meters
L