User manual

3
Before You Start
Safety Information
Overhead clearance is required. Do not place the instrument under a cabi net or other
structure. Always leave enough clearance to allow for safe and easy insertion and
removal of probes.
The instrument is intended for indoor use only.
DO NOT use the unit in environments other than those listed in the users guide.
DO NOT carry the unit without the transport lid in place and properly se cured.
Thoroughly wipe the inside of the well dry when changing uids as some high
temperature uids react violently to water or other liquid mediums.
Use of this instrument at HIGH TEMPERATURES for extended periods of time
requires caution.
Completely unattended high temperature operation in not recommended.
BURN HAZARD
High temperatures may be present in this equipment. Fires and severe burns may result
if personnel fail to observe safety precautions.
DO NOT mix water and oil when exceeding temperatures of 90°C (194°F).
ELECTRICAL HAZARD
These guidelines must be followed to ensure that the safety mechanisms in this
instrument will operate properly. This instrument must be plugged into an AC only
electric outlet as listed in Section 3.1, Specications. The power cord of the instrument
is equipped with a three-pronged grounding plug for your protection against electrical
shock haz ards. It must be plugged directly into a properly grounded three-prong
receptacle. The receptacle must be installed in accordance with local codes and
ordinances. DO NOT use an extension cord or adapter plug.
DO use a ground fault interrupt device. This unit contains a liquid. A ground fault
device is advised in case liquid is present in the electrical system and could cause an
electrical shock.
Always replace the power cord with an approved cord of the correct rating and type. If
you have questions, contact an Authorized Service Center (see Section 1.3).
High voltage is used in the operation of this equipment. Severe injury or death may
result if personnel fail to observe the safety precautions. Before working inside the
equipment, turn off the power and disconnect the power cord.
DO NOT connect this unit to a non-grounded, non-polarized outlet.
Always replace the fuse with one of the same rating, voltage, and type.