User manual
Fluke DAQ
Users Manual
The dialog box entries include:
Configuration file
Click Open… to load an existing configuration file.
Network Type
Select Isolated if the PC is connected directly to the
device, or General if the instrument is part of a network,
USB or mixed LAN and USB configuration.
Port Number
4369 is the default port number for the 2680 Series and
NetDAQ products. The default port for the 2638A Hydra
Series III and 1586A Super-DAQ is 3490. Fluke
recommends that you use the default port number.
Group is externally wired
2680A and NetDAQ devices may be wired together in a
master/slave arrangement.
Advanced Setting autodisable scan – Interval 3
(autoscan rate) (Not applicable for 2638A/1586A)
Interval 3 is the rate at which alarm conditions are
monitored. If disabled, alarm conditions are monitored
only when the associated channel is scanned during a
regular scan interval.
Max 999999 seconds
Min 0.001 seconds
3. To add a device, select the .cfg file in the TreeView
panel and click Insert.
4. Specify the Base Channel (1 to 99) and Device Type
and click OK. See the screen below. A device is
attached to the new configuration. As part of the
insertion process, a base channel number can be
specified. This number is to differentiate channels in
different devices. Insert a maximum of 99 Base
Channels of the same Device Type. The Base
Channel number auto-increments when devices are
added.
16