User guide

Fluid Management Customer Service 1.800.462.2466
14 | V-Series
Maintenance, Care, and Use
To ensure safe, dependable operation of your V-Series Mixer follow the maintenance
schedule detailed below:
Clean every day:
Clean all accessible inside and outside surfaces with a mild cleaning solution.
Clean control panel surfaces with a mild cleaning solution.
Every six months:
Level the machine.
Adjust belt tension as needed.
Check for dirt accumulations on motor. Remove dirt by vacuuming.
Every year:
Lubricate the sleeve bearings on the motor with 30-35 drops of SAE 20 oil.
Change the drive belt.