Error and Warning Messages
45
en
After pressing the start key: Red
and green tube status LEDs flash
slowly
Warning beep, 1x short
The pump was started without an
inserted tube or the tube was not
detected by the pump, e.g. defec-
tive transponder.
Insert a new tube into pump before starting.
Device information
Green tube status LED lights up Tube OK - irrigation can be started
All LEDs light up briefly and then
go out
Warning beep, 1x long
Self-test concluded successfully.
The pump is ready
Suction does not automatically
shut off when instrument is
closed
Leakage in tube system or defec-
tive vacuum suction
Check the tube system and suction container for
leaks. If there is no leakage, suction is defective and
must be switched off. It is possible to continue using
the device without integrated vacuum pump. In this
case, use external suction. After the surgical proce-
dure, have the device checked by an authorized ser-
vice technician.
Error and Warning Messages Cause Troubleshooting