11.4 Guidelines and Manufacturer's Statement/Electromagnetic In-
terference Immunity
The deviceFLUID CONTROL Lap 2216 is intended for use in an electromagnetic en-
vironment as described below. The user/operator of the device FLUID CONTROL
Lap 2216 must make sure the device is operated within such an environment.
*Note: U
T
is the mains alternating voltage before applying the test levels.
Electromagnetic
Interference Immu-
nity Tests
IEC 60601-1-2
test level
Concordance
Level
Electromagnetic environ-
ment/guidelines
Discharge of static
electricity (ESD)
according to
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be made
from wood or concrete or
covered with ceramic tiles.
If the floor covering con-
sists of synthetic material,
the relative humidity
should be at least 30%.
Electrical fast tran-
sients/bursts
according to
IEC 61000-4-4
± 2 kV for AC
power lines
±1kVfor input
and output lines
Modulation 100
KHz
± 2 kV for AC
power lines
±1kVfor
input and out-
put lines
Modulation
100 KHz
The quality of the supply
voltage should be the same
as the voltage of a typical
business or hospital envi-
ronment.
Voltage surges
according to
IEC 61000-4-5
± 1 kV line(s) to
line(s) ± 2 kV
line(s) to ground
± 1 kV line(s) to
line(s) ± 2 kV
line(s) to earth
The quality of the supply
voltage should be the same
as the voltage of a typical
business or hospital envi-
ronment.
Blackouts, brown-
outs, and fluctua-
tions of the power
supply according to
IEC 61000-4-11
< 5% U
T
* (> 95%
dip in the U
T
) for
½ cycle
< 5% U
T
*
(> 95% dip in
the UT) for 1/2
cycle 40% U
T
(60% dip in the
U
T
) for 5 cycles
70% U
T
(30%
dip in the U
T
)
for 25 cycles
<5%U
T
(> 95% dip in
the U
T
) for 5
seconds
0% U
T
;
½ cycle at 0°,
45°, 90°, 135°,
180°, 225°,
270°, and 315°
0% U
T
; 1 cycle
and 70% U
T
;
25/30 cycles
single phase:
at 0° 0% U
T
;
250/300 cycles
The quality of the supply
voltage should be the same
as the voltage of a typical
business or hospital envi-
ronment. If the user/opera-
tor of the system requires
the continuation of func-
tionality after power inter-
ruptions/disruptions, it is
recommended to supply
the device with power from
an uninterruptible power
supply.
< 40% U
T
(> 60%
dip in the U
T
) for
5 cycles
< 70% U
T
(> 30%
dip in the U
T
) for
25 cycles
< 5% U
T
(> 95%
dip in the U
T
) for
5 cycles
Supply frequency
magnetic field (50/
60 Hz) according to
IEC 61000-4-8
30 A/m 30 A/m Magnetic fields of the
mains power frequency
should comply with the
typical values of business
and hospital environments.