User Manual
Table Of Contents
38
Technical Data
en
1
A tube set is not an application part in terms of the standard. However, it meets
all the technical requirements for an application part.
Max. negative suction pressure [kPa] -60
Max. suction power [l/min] 2.0
Max. outlet pressure [mmHg] 400
Max. irrigation output [l/min] 2.0
Interfaces/ports:
Signal input/output Service interface (D-Sub female 9-pin
DE-9, only for service, no ESD protec-
tion)
Power supply IEC-60320-1 C14
RFID transponder technology
Transmit/Receive Frequency Range 13.56 MHz ± 0.424 MHz
Transceiver class Class I
RF Output Power -10.83 dBμA/m at 10 m/32.8 ft
Type of Antenna Inductive Loop Antenna
Antenna loop area 0.00032 m²
Modulation Amplitude-shift keying (ASK)
Mode of Operation (Simplex/Duplex) Duplex
Essential performance Irrigation flow generation (LAP) up to
2.0 l min, tolerance +/- 10%.
Generate suction pressure up to 60 kPa
for fluid suction