User Manual
Table Of Contents
Care and Maintenance
33
en
• Additional information
When sterilizing several tube sets, instruments, etc. with one sterilization cycle,
make sure not to exceed the max. load of the sterilizer.
Sterilization with ethylene oxide (ETO) is possible but has not been approved by
the manufacturer.
7.4 Annual Inspection
Manufacturer's specificationsThe manufacturer stipulates that qualified personnel or hospital technicians
must regularly test the device to assess its functionality. These tests must be per-
formed on an annual basis. The tests are described in chapter 8 "Annual Inspec-
tion".
Regular inspections will assist in early detection of possible malfunctions. This
helps preserve the device and increases its safety and service life.
NOTE!
Service and maintenance work may not be carried out during surgery.
7.5 Maintenance by Authorized Service Technician
Two-year maintenance intervalAn authorized service technician has to inspect and service the device at appro-
priate intervals to ensure the safety and functionality of the unit. The minimum
service interval is two years, depending on frequency and duration of use. If the
service interval is not maintained, the manufacturer does not assume any liabil-
ity for the functional safety of the device.
A sticker located on the rear panel of the device will remind you of the latest date
for the next service or maintenance check.
Authorized service technicians are trained and certified only by the manufactur-
er.
Authorized trained personnelAll of the service tasks, such as changes, modifications, repairs, calibrations, etc.
may be carried out only by the manufacturer or manufacturer-approved trained
and skilled technicians.
Unauthorized personnelThe manufacturer is not liable for the operational safety of the device if unau-
thorized persons conduct this maintenance or any other service tasks.
LiabilityUnauthorized opening of the device and repairs performed by unauthorized per-
sonnel or third parties and/or changes or modifications release the manufactur-
er of any liability concerning the operational safety of the device.
Technical documentsReceiving technical documentation from the manufacturer does not authorize
individuals to perform repairs, adjustments, or alterations on the device or acces-
sories/peripherals.
CertificationAsk the service technician for a certificate after he or she has inspected the unit
or performed any service tasks. This certificate lists the type and scope of the ser-
vice as well as the date and name of the servicing company together wit the sig-
nature of the service technician.