7 Care and Maintenance
Special care is necessary when servicing, maintaining, and storing the device and
associated accessories to maintain the functionality of both.
7.1 Cleaning the Device
1. Use the ON/OFF button to turn the device off.
2. Remove the power cable.
3. Wipe the surface of the device with a soft cloth moistened with the surface
disinfectant based on alcohol or aldehyde (for example Meliseptol® rapid).
The concentration of the used disinfectant depends on the information pro-
vided by the manufacturer of the disinfectant. Make sure moisture does not
enter the device.
NOTE!
Do not sterilize the device.
7.2 Care and Maintenance of Reusable Tube Set
The device can be used with reusable tube set validated for this purpose. Please
comply with the following information when using a reusable tube set.
Information about the pressure chamber
membrane
DANGER!
Remove the membrane before cleaning.
1. Before cleaning: Carefully remove the membrane (2) from the pressure
chamber (1). Make sure not to damage the membrane. This is done by pulling
the strap of the membrane (6) towards the top.
2. Reinstalling membranes before sterilization: Place lip (3) of membrane into
ring groove (1) of pressure chamber. The strap (6) must be positioned above
the provided notch (5).
3. Press lip of membrane into groove (4). A properly inserted membrane is flush
with the pressure chamber and exhibits no wrinkles.
7.2.1 Cleaning the Reusable Tube Set
1. Dismantle the tubing set.
2. Use a soft brush or cleaning gun with a pressure of 3 to 4 bar to flush and
rinse the individual parts.
3. Rinse individual parts with demineralized water.
4. Use compressed air or a sterile soft cloth to dry all parts.
7.2.2 Disinfecting the Reusable Tube Set
1. Only disinfect a thoroughly cleaned tube set.
2. Place all tube set components into an enzymatic (e.g. Neodisher® medizym)
or slightly alkaline (e.g. Neodisher® mediclean) disinfectant solution for a
short period. Do not stack parts. The concentration and the application dura-
tion of the disinfectant depends on the information provided by the manu-
facturer of the disinfectant. The tube set can be damaged if the concentration
is set too high.
3. Remove the parts from the solution using forceps with a soft edge.
4. Remove all disinfectant residues.
5. Make sure all disinfectant residue is removed from the individual parts.
6. Dry all parts with a sterile cloth and wrap each part in a separate sterile cloth.
7. Assemble all parts before sterilization.