5.8 Replacing the Suction Container
WARNING!
Full secretion containers must be replaced immediately without stopping sur-
gery. If the overflow protection of the secretion containers is triggered, suction
is stopped to prevent the ingress of fluids.
1. Stop the suction (see chapter 5.7 "Suction").
Press the suction start/stop key (LED goes out).
2. Actuate the trumpet valve or open the stopcock to vent the system.
3. Remove tubes from filled suction container.
4. Reattach tubes to empty suction container.
Close stopcock if using instrument with stopcock.
5. Press the suction start/stop key (LED lights up).
6. Starting suction:
Operate the trumpet valve or open the stopcock.
5.9 Removing a Vacuum Tube System
1. Remove the drain tube (4) from the instrument (5) (see Abb. 5-6 "Suction
system").
2. Remove the suction container (3) from the drain tube (4).
3. Remove the vacuum tube (1) from the suction container (3).
4. Remove the hygiene filter from the vacuum tube (1).
5. Remove the vacuum tube from the device.