User Manual

24
Operating the Device
en
Disposable tube set: After inserting the tube set and starting the irrigation cy-
cle, the tube set is invalidated. If the irrigation cycle is stopped, the device can
be restarted within 45 minutes. If the device is switched off or in case of a pow-
er failure, the tube set is invalidated. The irrigation cycle can no longer be start-
ed. If this is the case, you must insert a new, valid, and approved tube set.
If an invalid tube is inserted into the pump, the tube status LED lights up red
and three short warning beeps are emitted. The red LED flashes if irrigation is
still started. Irrigation can no longer be started.
Reusable tube set: The reusable tube set can be used about 20 times. After in-
serting and starting irrigation, one use cycle is invalidated. If irrigating is
stopped during the use cycle, irrigation can be restarted within 45 minutes. If
the device is switched off or in case of a power failure, the current use cycle is
invalidated. Once the last use cycle has been invalidated, irrigation can no lon-
ger be started. If this is the case, you must insert a new, valid, and approved
tube set.
If the tube set is used for the last time (20 times), the green tube status LED
flashes, i.e. the tube can no longer be used once the current procedure is fin-
ished.
NOTE!
Comply with hygiene rules and regulations when disposing of the tube set, col-
lected fluid, and the suction container.
Transponder signal loss The device stops and cannot be restarted if the transponder loses its signal dur-
ing current use. If the signal is restored with 60 seconds, it is possible to continue
using the tube set.
Diethylhexylphthalate (DEHP)
DANGER!
The vacuum tube sets for this device contain diethylhexylphthalate (DEHP),
which is classified as toxic to reproduction according to the EU Directive 1272/
2008/EEC on Classification, Labeling and Packaging of Dangerous Substances.
DEHP may impair fertility and may cause harm to the unborn child. Therefore,
this product must not be used for unauthorized applications. When applied
within the intended use, the potential risk to pregnant or breastfeeding women
as well as to children resulting from the DEHP contained in this product is not
critical.