User Manual
Table Of Contents
Operating the Device
23
en
A sterile technician then removes the inner tube set package and opens it.
Open sterile autoclavable container1b. Reusable tube set - To be carried out by non-sterile technician:
Open the sterile autoclavable container of the tube set.
A sterile technician then removes the tube set from the inside of the pack-
age.
Connect instrument2. Sterile tasks
Keep the Luer lock connector (10) in the sterile area and hand the tube end
with the tap spikes (6)/Care-Lock® (7) to the non-sterile personnel.
Remove the protective cap on the Luer lock connector.
Connect the Luer lock connector (10) with the instrument (e.g., inflow can-
nula). Open inflow valve at instrument.
Insert tube set3. To be carried out by non-sterile personnel:
Before inserting the tube, connect both tube clamps (5) (Abb. 5-2 "Tube set
irrigation elements") to the bags.
Switch the device on. The power status LED on the ON switch lights up. Make
sure all tube sets are removed from the pump before switching the device
on. A long beep is emitted to signal the successful completion of the device
test and the tube can now be inserted.
Inserting the roller tube is depicted in Abb. 5-4 "Inserting the roller tube".
When inserting the roller tube, make sure not to damage the membranes of
the pressure chamber. Insert the pressure chamber (2) only if chamber is
not pressurized.
Insert the unpressurized pressure chamber (2) carefully into the lower
notch of the tube retainer (12) up to the stop.
Place the roller tube (1) around the roller wheel.
Insert the tube adapter into the left section of the tube retainer by (13) tight-
ening the tube.
Make sure the tube adapter is firmly snapped into place! Irrigation and drain
tube must point straight to the front from the tube retainer.
Connect an instrument and the corresponding bags (see also chapter 5.5 "Con-
necting the Irrigation Fluid Bags").
WARNING!
Instrument tube (9) and irrigation tube (3) must be inserted vertically towards
the front without twisting the roller tube.
RFID technology (transponder technology)Transponder technology detects the tube type, the validity and reliability of a
tube set automatically. This eliminates virtually all "operating errors" since non
fitting, invalid, and not approved tube sets are reliably detected.
The tube status LED lights up green if a valid tube is inserted. Irrigation can be
started.
Invalidating a tube setIf an approved tube set is in the device, the transponder technology automatical-
ly invalidates this tube set after 60 seconds when the device is started. Depend-
ing on the tube set, this has the following consequences:
Abb. 5-4 Inserting the roller tube
(1) Roller tube
(2) Pressure chamber with membrane
(3) Irrigation tube
(9) Instrument tube
(12) Tube clamp
(13) Tube adapter
2
1
4
3
(2)
(13)
(1)
(3)
(12)
(9)