User Manual
Table Of Contents
Operating the Device
21
en
Before surgery1. Tube set and instrument are filled with irrigation fluid. The suction system is
connected. The stopcocks are closed.
2. Press the suction start/stop key (LED lights up).
The suction system is pre-evacuated.
Only then can the suction procedure be started with the instrument.
During surgery1. Press the irrigation start/stop key (LED lights up).
2. Insert instrument into abdomen.
Start irrigating wash:
Operate the trumpet valve or open the stopcock.
3. Starting the suction process:
Operate the trumpet valve or open the stopcock.
After surgery
NOTE!
The manufacturer recommends emptying the tube set using the pump when em-
ploying saccharine irrigation fluids (HF use). Remove irrigation fluid bags and in-
strument first.
1. Press the irrigation start/stop key and the suction start/stop key (LED goes
out).
2. Use the power switch to turn device off.
3. Remove the tube sets.
NOTE!
Comply with hygiene rules and regulations when disposing of the tube set, col-
lected fluid, and the suction container.
5.3 Overview of the Usable Tube Sets
The following table shows an overview FLUID CONTROL Lap 2216 of the tube sets
that can be used.
Tube category Art. No. Tube type
Irrigation tube set with tapping spike 8171223 Reusable (autoclavable)
Irrigation tube set with tapping spike L 3
m
4171223 Disposable (sterile)
Irrigation tube set with Care-Lock L 3 m 4171224 Disposable (sterile)
Suction tube set NL 3 m 4171225 Disposable (sterile)
Vacuum tube set 8170.401 Reusable (autoclavable)