User Manual

14
Device Startup
en
DANGER!
Set up the device in such a way as to allow for easy monitoring of the display val-
ues, device functions, and access to the control elements.
DANGER!
Position the device in such a way that it is easy to operate and switch off.
DANGER!
Risk of electrical shock
To avoid the risk of electrical shock, only use this device when connected to a
properly grounded power supply network.
DANGER!
Use only parts and/or devices from ME systems (see chapter 14 "Glossary") in pa-
tient environments in compliance with the standard IEC 60601-1 in the respec-
tive currently valid version
WARNING!
The ME device is suitable for arrangement in ME device systems (see chapter14
"Glossary"). Operation of the ME device in close proximity to non-ME devices
may result in the intended use of the ME device no longer being guaranteed.
WARNING!
Equipment should be positioned such that power cord can be easily disconnect-
ed.
WARNING!
Ventilation of the device
Avoid device overheating. Ensure free air circulation especially to the bottom
and rear of the device (rear panel distance of at least 10 cm).
WARNING!
Before switching on the device, sufficient time must have passed to adjust to the
indoor climate.
WARNING!
Place the device outside the sterile field.
WARNING!
Position of the user
To avoid a malfunction, the user must be positioned correctly
for observation up to 2 m from the front of the device.