en
DANGER!
The reusable tube set is made of silicone, polysulfone (PSU), and stainless steel.
For cleaning and disinfection, use only pH-neutral or slightly alkaline cleaning
agents (e.g. neodisher MediClean 2.0 %), disinfectants (e.g. Lysetol V 8 %), and
drying and rinsing agents approved for the tube set materials.
Using unsuitable agents (e.g. neodisher MediKlar rinsing agent) may damage
the tube system and especially the PSU connectors.
DANGER!
Always check the reusable tube set for signs of deterioration before use and after
sterilization. Never use a tube set which shows signs of deterioration, including
cracking, or perforation.
DANGER!
Make sure to connect tubes as indicated by the matching connection pieces (con-
nectors).
DANGER!
The day tube remains connected to the device after each surgery. The patient
tube must be disposed of immediately after surgery. A new sterile cap (included
with the patient tube) must be screwed onto the hygiene connector at the day
tube after each surgery. This sterile protector remains connected to the day tube
until the next surgery.
DANGER!
Do not touch patient and the ON/OFF switch at the same time.
DANGER!
Serum sodium concentration
It is also necessary to monitor the concentration of sodium in the blood of the pa-
tient to prevent electrolyte disturbances. Monitoring of the concentration of so-
dium in the blood must be performed by the physician and is not performed or
supported by the system.
DANGER!
Air Embolisms
An air embolism can be the result of air contained in the tube set or connected
instrument reaching the patient. Make sure that there is always fluid in the bag
to prevent air from being aspirated into the patient.