User Manual

General Information
9
en
DANGER!
Connect only items, that have been specified as part of the ME system or speci-
fied as being compatible with ME systems.
DANGER!
Please read the instruction manual for the stand before mounting the pump on
the stand.
DANGER!
The position of the device on the stand can affect the stand's stability. If the de-
vice is mounted too high on the stand, the stand may tilt or topple. The user/op-
erator of the device is responsible for making sure the device cannot be tipped
over or otherwise become unstable.
DANGER!
Please read the instruction manual for the device cart before mounting the
pump on the cart.
DANGER!
The device is only intended for use with flexible fluid containers. Do not use glass
containers as they might break. Fluid cannot flow quickly enough due to the vac-
uum being generated inside of the bottle. Risk of implosion.
DANGER!
Functional test
The functional test must be performed prior to each surgery.
DANGER!
Sterilize reusable instruments and tubing before surgery to prevent infections.
Check all the single-use/disposable items before removing them from the pack-
age to ensure that the packaging is intact and that the expiration date is still val-
id.
DANGER!
The irrigation fluid bags must not be stored on the device or hung up directly
above the device.
DANGER!
Select an irrigation fluid suitable for the subsequent medical procedure.
DANGER!
Checking the warning signals
The warning signals must be checked prior to each operation. The system is to be
set up so that all warning signals can be perceived.