Safety Notes

Error and Warning Messages
109
en
18 Error and Warning Messages
Error and warning messages are indicated by the device acoustically and visually.
The visual indication is usually a message in the status line of the touchscreen or
a full screen display. Acoustic messages are usually warning signals emitted a
certain number of times (beeps).
NOTE!
Each warning signal is to be assigned one of the corresponding priorities. If mul-
tiple signals of equal priority are called at the same time, they are processed ac-
cording to an internal priority ranking.
Error/error messages, full screen display
Warning messages depicted in status line
Error/error message Troubleshooting
Device error <-> Call service <->
Electronics error
2 warning beeps
Turn the device off and back on after approx.
10 seconds. If the malfunction or error mes-
sage is displayed again, the device may not
be used any longer. Make sure the device can
no longer be operated until an authorized
service technician conducts the appropriate
tests and repairs.
Touchscreen stays black
Device error <-> Call service <->
Sensor error
3 warning beeps
Device error <-> Call service <->
Motor error
5 warning beeps
Device error <-> Call service <->
Calibration error
10 warning beeps
Warning message Remedy
Overpressure!
Continuous acoustic signal
(beep)
Physician must reduce pressure
Critical overpressure!
Continuous acoustic signal
(beep)
Day tube contaminated
3 short warning beeps
Replace day tube
Vacuum pump defective!
3 long warning beeps
The suction function can no longer be used.
Make sure the device can no longer be oper-
ated until an authorized service technician
conducts the appropriate tests and repairs.
Deficit limit reached!
3 long warning beeps
Physician must respond appropriately
Deficit limit exceeded!
3 short warning beeps
Physician must respond appropriately
Perforation!
Continuous acoustic signal
(beep)
OBJ_DOKU-0000006143-008.fm Seite 109 Mittwoch, 11. November 2020 2:29 14