Safety Notes
Electromagnetic Compatibility
107
en
IMMUNITY TESTS
Transients HF IEC 61000-4-3
WARNING!
Portable HF communication devices can have an effect on the performance char-
acteristics of the device PG130/PG145. Therefore, such devices must be kept at
a minimum distance of 30 cm (independent of any calculation) from the device
PG130/PG145, its accessories, and the cables.
Test frequency
(MHz)
Band (MHz) Service Modulation Max. power
(W)
Distance (m) IMMUNITY
TEST
LEVEL
(V/m)
385 380-390 TETRA 400 Pulse modula-
tion
18 Hz
1.8 0.3 27
450 430-470 GMRS 460, FRS
460
FM
±5 kHz devia-
tion
20.328
710
704-787 LTE Band 13, 17
Pulse modula-
tion
217 Hz
0.2 0.3 9
745
780
810
800-960
GSM 800/900,
TETRA 800,
IDEN 820,
CDMA 850, LTE
Band 5
Pulse modula-
tion
18 Hz
20.328
870
930
1720
1700-1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE Band
1, 3, 4, 25;
UMTS
Pulse modula-
tion
217 Hz
20.328
1845
1970
2450 2400-2570 Bluetooth,
WLAN,
802.11 b/g/n
RFID 2450, LTE
Band 7
Pulse modula-
tion
217 Hz
20.328
5240
5100-5800
WLAN 802.11
a/n
Pulse modula-
tion
217 Hz
0.2 0.3 9
5500
5785
OBJ_DOKU-0000006143-008.fm Seite 107 Mittwoch, 11. November 2020 2:29 14