Electromagnetic Compatibility
en
17 Electromagnetic Compatibility
Precautionary measures Medical devices are subject to special safety and protective measures concerning
electromagnetic compatibility (hereafter abbreviated as EMC).
This device is to be used only for the purposes described in the manual and has
to be installed, set up, and operated in compliance with the EMC notes and in-
structions.
CAUTION!
Accessories
To ensure compliance with the requirements of IEC 60601-1-2 in the current ver-
sion, the device PG130/PG145 must be used only with the accessories listed in
Chapter 20 Article/Accessory List PG130/PG145.
To ensure the basic safety and essential functionality concerning electromagnet-
ic interference over the life of the device, the device must be restarted no later
than after 24 hours so that a diagnostic self-test can be performed. In addition,
the specified maintenance intervals must be observed (see Chapter 15.2 "Main-
tenance Intervals").
17.1 Electrical Connections
ESD (Electrostatic Discharge) precautionary
measures
The following are ESD precautionary measures:
Apply potential equalization (PE), if available on your equipment, to all devices
to be connected.
Use only the listed equipment and accessories.
Hospital employees should be informed about and trained in ESD precautionary
measures.
17.2 Guidelines and Manufacturer’s Statement – Electromagnetic
Emissions
The PG130/PG145 is intended for use in an environment as described below. The
user/operator of the PG130/PG145 should make sure the device is operated
within such an environment.
Emitted interference
measurements
Compliance Electromagnetic environment guide-
lines
HF emission according
to CISPR 11
Group 1 The PG130/PG145 uses HF energy
solely for its internal functions.
Therefore, the camera's HF emission
is very low and it is unlikely that
devices in close proximity will experi-
ence interference.
HF emission according
to CISPR 11
Class B The PG130/PG145 is suitable for use
in all facilities including those in resi-
dential areas and those directly con-
nected to a public utility network
supplying buildings used for residen-
tial purposes as well.
Emission of harmonic
oscillations according
to IEC 61000-3-2
Class A
Emission of voltage
fluctuations / flickers
according to
IEC 61000-3-3
In compliance