Safety Notes
Annual Inspection
103
en
sensor.
9. The actual pressure display should depict approx. 5 mmHg.
10. Hold the instrument tube at a level of approx. 67.5 cm above the roller wheel.
Make sure the tube segment between this point and the cartridge is com-
pletely filled with fluid. Also make sure that no point of the tube is higher
than the indicated value.
11. The height of the water column is used to calculate the generated hydrostatic
(water) pressure according to 67.5 cm H
2
O x 0.74 = 50 mmHg.
12. The actual pressure display now should depict approx. 50 mmHg
(±10 mmHg).
13. When varying the height of the tube set above the roller wheel, the actual
pressure display also varies accordingly.
The test of the pressure measurement is successfully completed when the values
of the actual pressure display correspond to those of the converted water column
height. Please enter each completed test into the test log (see Chapter 21.1 "Test
Log").