Safety Notes
Annual Inspection
101
en
The layout of the test and its setup are depicted in Fig. 16-1 "Setup of basic func-
tion test".
1. Switch device on. Wait until the device check has finished.
2. Select one of the possible indications. If you have selected the Arthroscopy in-
dication, then press the [Knee] function field.
3. Insert a disposable tube set into the tube retainer.
4. Suspend fluid bag at a height illustrated in Fig. 16-1 "Setup of basic function
test" and connect bag with irrigation tube.
5. Place the end with the luer lock into the measuring cup.
6. Set the following nominal values:
Arthroscopy, hysteroscopy, and urology indications:
• Nominal pressure = 90 mmHg
• Nominal flow = 500 ml/min
Laparoscopy indication:
• Flow standard = 1 l/min
7. Press the [START] function field.
8. Wait until the tube set is completely filled with irrigation fluid.
Arthroscopy and Urology Indications
Allow the device to pump fluid for at least 1 minute (instrument recognition
function).
9. Clamp off the instrument tube end in the measuring cup without stopping
the pump.
10. Empty the measuring cup and place the end of the tube back into the mea-
suring cup.
11. Release the tube end. The irrigation process runs. Press the start button of the
stopwatch.
12. Arthroscopy, hysteroscopy, and urology indications
After 2 minutes, press the [STOP] function field and the stop key of the stop-
watch at the same time. The measuring cup then contains approx. 1 l (±10 %)
Fig. 16-1 Setup of basic function test
h=x cm
!