Safety Notes

98
Care and Maintenance
en
5. A Undo the latch of the fuse holder with a small screwdriver.
6. B Remove the fuse holder.
7. C Check fuses.
8. Insert a new fuse. Use only the specified type of fuse (see Chapter 19 Techni-
cal Data).
9. Insert the fuse holder until it can be heard snapping into place.
10. Use the power cable to reconnect the shockproof safety socket with the rear
mains socket.
15.6 Cleaning the Accessories
1. Switch device off.
2. Disconnect device from power supply.
3. Disconnect the accessories (foot pedal and scale) from the device.
The concentration of the used disinfectant depends on the information provided
by the manufacturer of the disinfectant. Wipe with a soft cloth moistened with
the disinfectant. The ingress of liquid into the accessories such as foot pedal,
scale, and remote control must be prevented at all costs.
The manufacturer recommends Meliseptol rapid from B. Braun as a disinfectant
for the foot pedal and scale. The manufacturer recommends Incidin Extra N from
ECOLAB Healthcare as disinfectant for the remote control.
CAUTION!
Do not use any alcohol-based desinfectants on the remote control.
WARNING!
Cleaning the Accessories
The accessories (foot pedal, scale, remote control) may not be sterilized.
15.7 Care of the Reusable Tube Set
WARNING!
Please check the reusable tube for signs of damage after sterilization and before
use. Never use a tube showing signs of damage, especially brittleness and perfo-
ration.
CAUTION!
The reusable tube system consists primarily of silicone, polysulfone (PSU), and
stainless steel. For cleaning and disinfection, use pH-neutral or slightly alkaline
cleaners and disinfectants approved for the utilized materials. If using unsuit-
able cleaners or disinfectants, the tube system and especially the PSU connectors
may become damaged.
CAUTION!
Recycling limitations
The manufacturer has inspected the reusable tube system for a limited number
of recycling cycles. Always comply with the instructions on the label. Never ex-
ceed the number of uses indicated by the manufacturer.
Application site Use a soft cloth immediately after surgery to remove tissue, blood, etc. complete-
ly. This cloth may be moistened with a slightly alkaline cleaner.
Sticky coagulates can be removed carefully using a brush.
OBJ_DOKU-0000006143-008.fm Seite 98 Mittwoch, 11. November 2020 2:29 14