Safety Notes
Care and Maintenance
97
en
15.4 Aesculap Technical Service
Technical service address
Please contact the address above for additional service addresses.
15.5 Replacing the Fuse
CAUTION!
Before replacing the fuse, check the values of the fuse to be inserted acc. to
Chapter 19 Technical Data.
The fuse may be defective and is in need of replacement if:
• displays and LEDs do not light up,
• the device does not function.
Check to make sure
• the main power supply cable is properly connected to the power supply input
and to a safety socket,
• the house power supply fuse is functioning.
WARNING!
Unplug the power cable from the device before checking the fuse.
The device does not have to be opened to replace the fuse.
1. Switch device off.
2. Disconnect device from power supply.
3. Remove power connection cable from mains socket.
4. The fuse holder is located to the left of the mains socket. Remove fuse holder
as depicted in Fig. 15-1.
Aesculap Technical Service
Am Aesculap-Platz
78532 Tuttlingen/Germany
Phone: +49 7461 95-1601
Fax: +49 7461 14-939
Email: ats@aesculap.de
Fig. 15-1 Opening the fuse holder
A
C
B
OBJ_DOKU-0000006143-008.fm Seite 97 Mittwoch, 11. November 2020 2:29 14