15 Care and Maintenance
Special care is necessary when servicing, maintaining, and storing the device and
its accessories to maintain the functionality of the device and its accessories.
15.1 Cleaning the Device
1. Use the ON/OFF button to turn the device off.
2. Remove the power cable.
3. Wipe the surface of the device with a soft cloth moistened with the surface
disinfectant based on alcohol (for example Meliseptol® rapid). The concentra-
tion of the used disinfectant depends on the information provided by the
manufacturer of the disinfectant. Make sure moisture does not enter the de-
vice.
NOTE!
Do not sterilize the device.
15.2 Maintenance Intervals
Manufacturer's specifications The manufacturer stipulates that qualified personnel or hospital technicians
must regularly test the device to assess its functionality and technical safety. The
device needs to be serviced at least every two years. The tests are described in
Chapter 16 "Annual Inspection".
Regular inspections will assist in early detection of possible malfunctions. This
helps preserve the device and increases its safety and service life.
NOTE!
Service or maintenance work may not be carried out during surgery.
15.3 Maintenance Carried out by Authorized Service Technician
Two-year maintenance interval An authorized service technician must inspect and service the device at appropri-
ate intervals to ensure the safety and functionality of the device. The minimum
service interval is two years, depending on frequency and duration of use. If the
service interval is not maintained, the manufacturer does not assume any liabil-
ity for the functional safety of the device.
A sticker located on the rear panel of the device will remind you of the latest date
for the next service or maintenance check.
Authorized service technicians are trained and certified only by the manufactur-
er.
Authorized trained personnel All services, such as modifications, repairs, calibrations, replacement of the re-
mote control battery, etc., may only be performed by the manufacturer or by spe-
cialists authorized by the manufacturer.
Unauthorized personnel The manufacturer is not liable for the operational safety of the device if unau-
thorized persons conduct this maintenance or any other service tasks.
Liability Unauthorized opening of the device and repairs performed by unauthorized per-
sonnel or third parties and/or changes or modifications release the manufactur-
er of any liability concerning the operational safety of the device.
Technical documents Receiving technical documentation from the manufacturer does not authorize
individuals to perform repairs, adjustments, or alterations on the device or acces-
sories/peripherals.
Certification Ask the service technician for a certificate after he or she has inspected the unit
or performed any service tasks. This certificate lists the type and scope of the ser-
vice as well as the date and name of the servicing company together with the sig-
nature of the service technician.