Safety Notes

Functional Test
95
en
"Aesculap Technical Service").
7. Exit the user menu by pressing [Exit].
8. Remove weight from scale.
The function test of the scale is successfully completed when the permissible tol-
erance of < 100 g is maintained.
14.6 Function Control of Foot Pedal (Only for Indication Arthroscopy)
1. A foot pedal is connected to the switched-on device.
2. Press the foot pedal to test the functions described in Chapter 6.15 Using a
Foot Pedal (Optional).
14.7 Function Check of the Remote Control
1. An optional remote control is connected to the switched-on device.
2. Press the remote control to test the functions described in Chapter 6.13 Using
the Remote Control (Optional).
14.8 Ending the Function Test
The device is ready for surgical use after the function test has concluded success-
fully. Position the device as needed for surgery.
WARNING!
The physician must determine an irrigation fluid suitable for the application and
medical procedure!
Please read Chapter 13 "Safety Functions" thoroughly before using the device.
This Chapter describes all safety features. The function test must be completed
successfully before each surgery.
1. Before surgery:
Select the desired nominal pressure and the nominal flow at the device.
Press the [START] function field. The roller wheel starts to turn.
Insert the instrument filled completely with irrigation fluid into the patient.
2. After surgery:
Close the inflow valve of the working shaft.
Press the [STOP] function field. The roller wheel stops.
Switch device off.
Remove the tube set and the fluid bags.
NOTE!
Observe applicable hygiene regulations when disposing of the tube set.
OBJ_DOKU-0000006143-008.fm Seite 95 Mittwoch, 11. November 2020 2:29 14