Safety Notes
94
Functional Test
en
Open the tube clamp and the inflow valve.
Press the [START] function field.
The roller wheel starts to turn. Wait until tube set and instrument are com-
pletely filled with water.
Press the [STOP] function field. The roller wheel stops.
Check user menu
Check the settings in the user menu (see Chapter 12.2 "General Settings").
14.2 Performing the Device Function Test (Only for Laparoscopy Indi-
cation)
1. Set the "Standard" flow rate.
2. Press the [START] function field.
3. Open inflow valve at instrument. The roller wheel is turning forward and de-
livers irrigation fluid.
4. Close the inflow valve of the instrument. The roller wheel stops.
5. Press the [STOP] function field. The roller wheel turns approx. one rotation
backwards.
The function test of the device is successfully completed when the roller wheel
behaves as described in item 3., 4., and 5.
14.3 Performing Device Function Test (Only for Arthroscopy, Hyster-
oscopy, and Urology Indication)
1. Set the nominal pressure to 50 mmHg and the nominal flow to 0.5 l/min.
2. Press the [START] function field.
3. Close the inflow valve of the instrument about halfway so that device can still
deliver fluid through the instrument.
The function test of the device is successfully completed when the progress bar
of the actual pressure depicts a value of approx. 50 mmHg.
14.4 Performing Function Test of the Wash Function (Only for Ar-
throscopy Indication)
1. Switch device on.
2. Press the [ARTHRO] function field. The joint selection is displayed (see Chap-
ter 9.5 "Selecting Joint Type").
3. Press the desired function field, e.g., [Knee].
4. Press the [START] function field.
5. Open inflow valve at instrument.
6. Press the [WASH] function field to enable the Wash function.
7. The Wash function is running.
8. The nominal values of the Wash function are displayed.
9. Press the [WASH] function field to end the Wash function early. The previous-
ly set nominal values are displayed.
14.5 Performing Scale Function Test (Only for Hysteroscopy Indica-
tion)
Test connected scale
1. A scale without weight is connected to the switched on device (see Chapter
10.6.1 "Installing and Connecting the Scale").
2. In the menu, press the [HYSTEROSCOPY] function field. This opens the hyster-
oscopy settings.
3. Press the [Scales test] function field.
4. The weight display depicts 0 g.
5. Place a defined weight between 1 to 5 kg (e.g., 3 kg) on the scale.
6. The weight display of the device must depict the applied weight in grams
(e.g., 3000 g). The permissible tolerance is < 100 g. If a greater difference is
detected, a service technician must re-calibrate the scale (see Chapter 15.4
Menu
Scale test
3000 g
7.