Safety Notes
User Menu
91
en
3. Press the [User menu] symbol. The user menu opens.
4. Press the [Service] function field.
5. Press one of the 4 function fields: [Software version], [Pressure sensor test],
[Scale test], or [Service menu].
12.3.1 Software Version
1. In the user menu, press the [Service] function field.
2. Press the [Software version] function field to open the corresponding display.
12.3.2 Performing a Pressure Sensor Test
The pressure sensor test is part of the annual inspection, see Chapter 16.3 "Pres-
sure Measuring Test".
12.3.3 Performing a Scale Test
Test connected scale1. A scale without weight is connected to the switched on device (see Chapter
10.6.1 "Installing and Connecting the Scale").
2. In the user menu, press the [Service] function field.
3. Press the [Scales test] function field.
4. The weight display depicts 0 g.
5. Place a defined weight between 1 to 5 kg (e.g., 3 kg) on the scale.
6. The weight display of the device must depict the applied weight in grams
(e.g., 3000 g). The permissible tolerance is < 100 g. If a greater difference is
detected, a service technician must re-calibrate the scale (see Chapter 15.4
"Aesculap Technical Service").
You can now do the following:
7. Press [o] to return to the previous menu level.
8. Press [Exit] to return to the screen display of the hysteroscopy indication.
The test of the scale is successfully completed when the permissible tolerance of
< 100 g is maintained.
12.3.4 Service Menu
Access to the service menu is restricted to trained and authorized service person-
nel. Please consult the service manual for further information.
ACTUAL PRESSURE
100 150050
SET
PRESSURE
80 mmHg
SET
FLOW
STOP
HYS
HIHIHI
LO
0
SET
PRESSURE
45 mmHg
SET
FLOW
HIHIHI
LO
ACTUAL PRESSURE
100 150050
INFLOW VOLUME
1.0
INFLOW VOLUME
52
INFLOW VOLUME
1.0
DEFICIT [ml]
0
100
%
0
0
STOPSTART
Start pump
HYS
100 ml/min
3.
General settings
HYS settings
Service
Menu
4.
Software version
Pressure sensor test
Scale test
Menu
Service
Service menu
Menu
Software version
XP2 011 DD YYYY
Menu
Scale test
3000 g
7.
8.
OBJ_DOKU-0000006143-008.fm Seite 91 Mittwoch, 11. November 2020 2:29 14