Safety Notes

84
Using the Urology Indication
en
pressure display shows the current measured value. The roller wheel starts to
turn and the device begins to run the instrument detection function.
The [START] function field can also be controlled with the help of the remote
control (see Chapter 6.13 "Using the Remote Control (Optional)").
6. The status line depicts Tube filling. Wait until the tube set is completely filled
with fluid.
7. Once the tube filling phase is finished, the status line depicts Tube filling
completed and three beeps are emitted.
8. Close the inflow valve of the instrument.
9. You may now start the surgical procedure.
The tube filling phase is restarted and the indicated in the status line after
each pump start
11.10 Stopping Irrigation
1. Press the [STOP] function field.
The [STOP] function field can also be controlled with the help of the remote
control (see Chapter 6.13 Using the Remote Control (Optional)).
2. The touchscreen then depicts the following values:
Actual pressure display: Current measured value
Nominal flow display: Last set value
Nominal pressure display: Last set value
11.11 Automatic Instrument Recognition
Improved pressure measuring due to the in-
strument detection function
The device features the automatic instrument detection function when using the
Urology. The automatic instrument detection function runs in the back after
each start of the irrigation process and serves to compensate pressure losses due
to the flow through the narrow working channel. The pressure drop at the instru-
ment is here included in the measured value of the actual pressure.
URO
SET
PRESSURE
45 mmHg
SET
FLOW
0.3 l/min
ACTUAL PRESSURE
0 50 100 150
Start pump
START
URO
SET
PRESSURE
45 mmHg
SET
FLOW
0.3 l/min
ACTUAL PRESSURE
0 50 100 150