Safety Notes
Using the Urology Indication
83
en
the values in increments of 10 mmHg.
The nominal pressure can also be reduced with the help of the remote control
(see Chapter 6.13 "Using the Remote Control (Optional)").
11.7 Preselecting the Nominal Flow
CAUTION!
The nominal pressure cannot be reached if the nominal flow is set too low.
The nominal flow can be increased or decreased while the device is being used or
not used. Values may be in the range of 0.025 to 0.5 l/min.
•Press the [+] symbol briefly to increase the nominal flow in increments of
0.025 l/min. Pressing the [+] symbol longer than 1.5 seconds activates scrolling
of the values in increments of 0.050 l/min.
The nominal flow can also be raised with the help of the remote control (see
Chapter 6.13 "Using the Remote Control (Optional)").
Factory setting: 0.25 l/min
• Briefly press the [-] symbol to decrease the nominal pressure in increments of
0.025 l/min. Pressing the [-] symbol longer than 1.5 seconds activates scrolling
of the values in increments of 0.050 l/min.
The nominal flow can also be reduced with the help of the remote control (see
Chapter 6.13 "Using the Remote Control (Optional)").
11.8 Saving Nominal Values
Save nominal valuesThe two set nominal values can now be saved. Next time the urethroscopy indi-
cation is opened, the last saved values for the nominal pressure and the nominal
flow are used automatically. If the nominal pressure value to be saved exceeds
45 mmHg, the nominal pressure value is reset automatically to 45 mmHg next
time the URO indication is opened.
•Press [Save] to save the entered nominal values.
11.9 Starting Irrigation
1. Insert the tube set as described in Chapter 8.4 "Inserting a Tube Set".
2. Open tube clamps at irrigation tube.
3. Open inflow valve at instrument completely.
4. If present, completely close the instrument outflow valve.
5. Press the [START] function field to start the irrigation process. The actual
URO
SET
PRESSURE
45 mmHg
SET
FLOW
0.3 l/min
ACTUAL PRESSURE
0 50 100 150
Start pump
START
URO
SET
PRESSURE
45 mmHg
SET
FLOW
0.3 l/min
ACTUAL PRESSURE
0 50 100 150
Start pump
START
URO
SET
PRESSURE
45 mmHg
SET
FLOW
0.3 l/min
ACTUAL PRESSURE
0 50 100 150
Start pump
START
URO
SET
PRESSURE
45 mmHg
SET
FLOW
0.3 l/min
ACTUAL PRESSURE
0 50 100 150
Start pump
START
OBJ_DOKU-0000006143-008.fm Seite 83 Mittwoch, 11. November 2020 2:29 14