Safety Notes
82
Using the Urology Indication
en
11.5 Switching Device on When Using Standard Tube Set
Switching device on without standard tube
set
1. Make sure no tube set is inserted into the tube retainer before switching the
device on.
2. Press the ON switch. The device turns on. The touchscreen displays the fol-
lowing in succession:
Company logo/Device check/Device OK
3. After the successful device check, the available indications are displayed for
selection (see also Chapter 6.3 "Activating a New Indication"). If the device
check is unsuccessful, please read Chapter 6.1.1 "Device Displays after Unsuc-
cessful Completion of Device Check").
4. Press the [URO] function field.
5. The Urology indication screen is displayed. The displayed parameters corre-
spond with the factory settings or the last saved nominal values. The status
line depicts the following: Insert tube set.
Tube set not valid In case an invalid tube set has been inserted, the status line alternately shows the
following: Tube set not valid and Please replace. A short acoustic alarm is emitted
3x.
11.6 Preselecting the Nominal Pressure
The nominal pressure can be increased or decreased while the device is being
used or not used. Values may be in the range of 15 to 90 mmHg.
Increasing nominal pressure • Briefly press the [+] symbol to increase the nominal pressure in increments of
5 mmHg. Pressing the [+] symbol longer than 1.5 seconds activates scrolling of
the values in increments of 10 mmHg.
The nominal pressure can also be raised with the help of the remote control
(see Chapter 6.13 "Using the Remote Control (Optional)").
Factory setting: 30 mmHg
When setting a nominal pressure value higher than 45 mmHg, a warning beep
is emitted 1 time and the status line depicts. >45 mmHg?.
Decreasing nominal pressure • Briefly press the [-] symbol to decrease the nominal pressure in increments of
5 mmHg. Pressing the [-] symbol longer than 1.5 seconds activates scrolling of
Device check
LAP
ARTHRO
HYS URO
4.
URO
SET
PRESSURE
45 mmHg
SET
FLOW
0.3 l/min
ACTUAL PRESSURE
0 50 100 150
Insert tube set
START
URO
SET
PRESSURE
45 mmHg
SET
FLOW
0.3 l/min
ACTUAL PRESSURE
0 50 100 150
Tube set not valid
START
URO
SET
PRESSURE
45 mmHg
SET
FLOW
0.3 l/min
ACTUAL PRESSURE
0 50 100 150
Start pump
START