Safety Notes
Using the Urology Indication
81
en
11.3 Screen Displays of the Urology Indication
11.4 Inserting a Tube Set
Tube types UrologyThe following tube sets can be used for the indication Urology (see also Chapter
20 Article/Accessory List PG130/PG145):
Table 11-1
A standard tube set is not inserted into the tube retainer until the device has
been switched on and the device check has been completed.
CAUTION!
When inserting a tube set already filled with irrigation fluid into the tube retain-
er, it is possible to damage the membranes. Only insert unfilled tube sets to
avoid such damage.
For additional information, see 6.5 "Using the Tube Sets", 6.7 "Inserting a Stan-
dard Tube Set" and 6.9 "Removing a Tube Set".
Fig. 11-1 Screen displays of the urology
indication
(1) Indication
(2) Actual pressure display
(3) Nominal flow display
(4) Suction function display (only
PG145)
(5) [+] symbol to increase nominal
flow
(6) [Indication selection] symbol
(7) [User menu] symbol
(8) [Save] symbol
(9) [Fluid bag] symbol
(10) [-] symbol to decrease nominal
flow
(11) Status line for messages
(12) [+] symbol to increase nominal
pressure
(13) [START]/[STOP] function field
(14) [-] symbol to decrease nominal
pressure
(15) Nominal pressure display
URO
SET
PRESSURE
45 mmHg
SET
FLOW
0.3 l/min
ACTUAL PRESSURE
0 50 100 150
Start pump
START
0
2
(1) (2) (3)
(4)
(6)(7)(8)(9)(10)
(14)
(15)
(5)
(13)
(12) (11)
Tube type Tube category
Disposable (one-
time use)
Disposable tube set, Luer lock connector, and tap spikes
Disposable tube set, Luer lock connector, and Care lock
Reusable Reusable tube set, Luer lock connector, and tap spikes
OBJ_DOKU-0000006143-008.fm Seite 81 Mittwoch, 11. November 2020 2:29 14