Safety Notes
Using the Hysteroscopy Indication
75
en
Table 10-2
Reset to factory settings1. Press the [Factory settings] function field to reset the HYS device parameters
to the factory settings.
You can now do the following:
2. Press [Save] to save the setting. After saving, the previous menu level is auto-
matically opened.
3. Press [o] to return to the previous menu level without resetting to the fac-
tory settings.
4. Press [Exit user menu] to return to the screen display of the hysteroscopy in-
dication without saving.
10.14 Safety Functions
10.14.1General Safety Functions
During device operation, the function display depicts safety and warning mes-
sages. These messages refer to the handling and safety of the device during use.
Please read Chapter 13 Safety Functions for additional information concerning
the general safety functions.
10.14.2Exceeding the Deficit Rate
Deficit rate exceededIf the deficit rate of 300 ml/min is being exceeded, a continuous warning beep is
emitted and the screen displays the following in red letters: Perforation!.
10.14.3Reaching and Exceeding the Deficit Threshold
The deficit threshold defines the threshold value of the scale for triggering the
warning messages (see Chapter 10.13.3 "Setting the Deficit Threshold").
Differential volume = threshold valueIf the differential (deficit) volume reaches the set threshold (100-3000 ml), a long
warning beep is emitted 3 times and the screen depicts the following: Deficit
reached!.
Hysteroscopy device parame-
ters
Hysteroscopy factory settings
Nominal pressure 45 mmHg
Flow level LO 100 ml/min
Flow level HI 300 ml/min
Deficit threshold 2000 ml
Flow level LO
Flow level HI
Deficit threshold
Menu
HYS settings
Factory settings
1.
Menu
Factory settings
Press Save button to reset all
indication presets to factory defaults
according to manual
4.
2.
3.
ACTUAL PRESSURE
100 150050
SET
PRESSURE
80
mmHg
SET
FLOW
STOP
HYS
HIHIHI
LO
0
SET
PRESSURE
45 mmHg
SET
FLOW
HYS
HIHIHI
LO
0
ACTUAL PRESSURE
100 150050
INFLOW VOLUME
1.0
INFLOW VOLUME
52
INFLOW VOLUME
1.0
DEFICIT [ml]
260
100
%
0
STOP
Perforation!
100 ml/min
ACTUAL PRESSURE
100 150050
SET
PRESSURE
80 mmHg
SET
FLOW
INFLOW VOLUME
1.0
HYS
HIHIHI
LO
SET
PRESSURE
45 mmHg
SET
FLOW
HYS
HIHIHI
LO
ACTUAL PRESSURE
100 150050
Deficit reached!
STOP
STOP
0
INFLOW VOLUME
1.0
DEFICIT [ml]
1000
100
%
0
100 ml/min
OBJ_DOKU-0000006143-008.fm Seite 75 Mittwoch, 11. November 2020 2:29 14