Safety Notes

74
Using the Hysteroscopy Indication
en
10.13.1Setting the Nominal Flow Level LO
Set nominal flow level LO 1. In the user menu, press the [HYS settings] function field. This opens the hys-
teroscopy user menu.
2. Press the [Flow level LO] function field. The nominal flow level LO can be set
in a range of 50-200 ml/min.
3. Press the [+] or [-] symbol to increase or decrease the nominal flow level LO.
Pressing this symbol briefly increases or decreases the flow in increments of
5 ml/min, pressing it longer in increments of 20 ml/min.
You can now do the following:
4. Press [Save] to save the setting. After saving, the previous menu level is auto-
matically opened.
5. Press [o] to return to the previous menu level without saving.
6. Press [Exit user menu] to return to the screen display of the hysteroscopy in-
dication without saving.
Factory setting: 100 ml/min
10.13.2Setting the Nominal Flow Level HI
Set nominal flow level HI 1. In the user menu, press the [HYS settings] function field. This opens the hys-
teroscopy user menu.
2. Press the [Flow level HI] function field. The nominal flow level HI can be set in
a range of 200-500 ml/min.
3. Press the [+] or [-] symbol to increase or decrease the nominal flow level HI.
Pressing this symbol briefly increases or decreases the flow in increments of
10 ml/min, pressing it longer in increments of 20 ml/min.
You can now do the following:
4. Press [Save] to save the setting. After saving, the previous menu level is auto-
matically opened.
5. Press [o] to return to the previous menu level without saving.
6. Press [Exit user menu] to return to the screen display of the hysteroscopy in-
dication without saving.
Factory setting: 300 ml/min
10.13.3Setting the Deficit Threshold
Set deficit threshold The deficit threshold defines the threshold valve for triggering the warning mes-
sages.
1. In the user menu, press the [HYS settings] function field. This opens the hys-
teroscopy user menu.
2. Press the [Deficit threshold] function field. The deficit threshold can be set
within a range of 100<a>3000 ml/min.
3. Press the [-] or [+] symbol to increase or decrease the deficit threshold. Press-
ing this symbol briefly increases or decreases the flow in increments of 10 ml/
min, pressing it longer in increments of 100 ml/min.
You can now do the following:
4. Press [Save] to save the setting. After saving, the previous menu level is auto-
matically opened.
5. Press [o] to return to the previous menu level without saving.
6. Press [Exit user menu] to return to the screen display of the hysteroscopy in-
dication without saving.
Factory setting: 2000 ml
10.13.4Resetting to Factory Settings
Use the hysteroscopy user menu to reset the hysteroscopy device parameters to
the factory settings. Please consult the following table for the device parameters
and the corresponding values:
Menu
Flow level LO
100 ml/min
3.
3.
6.4.
5.
Menu
Flow level HI
300 ml/min
3.
3.
6.
4.
5.
Menu
Deficit threshold
2000 ml
3.
3.
6.4.
5.