Safety Notes

Using the Hysteroscopy Indication
71
en
10.9 Resetting the Differential Volume
Reset differential volumeYou can reset the differential (deficit) volume to the value 0.
•Press the [->0<-] symbol (reset) to reset the differential (deficit) volume to the
value 0. The status line depicts the following: Deficit reset.
10.10 Starting Irrigation
1. Insert the tube set as described in Chapter 8.4 "Inserting a Tube Set".
2. Open tube clamps at irrigation tube.
3. Open inflow valve at instrument.
4. If present, completely close the instrument outflow valve.
5. Press the [START] function field to start the irrigation process. The actual
pressure display shows the current measured value. The roller wheel starts to
turn.
The [START] function field can also be controlled with the help of the remote
control (see Chapter 6.13 "Using the Remote Control (Optional)").
6. The status line depicts Tube filling. Wait until the tube set is completely filled
with fluid.
7. Once the tube filling phase is finished, the status line depicts Tube filling
completed and three beeps are emitted.
8. Close the inflow valve of the instrument.
9. You may now start the surgical procedure.
The tube filling phase is restarted and the indicated in the status line after
each ON/Off of the pump
•each removal of the tube during surgical mode
each tube change
each indication change
10.11 Stopping Irrigation
1. Press the [STOP] function field.
The [STOP] function field can also be controlled with the help of the remote
control (see Chapter 6.13 Using the Remote Control (Optional)).
2. The touchscreen then depicts the following values:
Actual pressure display: Current measured value
Nominal flow display: Last set value
Nominal pressure display: Last set value
10.12 Replacing the Secretion Container
CAUTION!
Full secretion containers must be replaced immediately without stopping sur-
gery. If the overflow protection of the secretion containers is triggered, suction
is stopped to prevent the ingress of fluids.
Container change during surgeryThe full secretion collection container can be replaced with an empty one during
surgery without losing the current measured value of the deficit volume.
1. Press the [STOP] function field.
2. Open inflow valve at instrument to vent the system.
ACTUAL PRESSURE
100 150050
SET
PRESSURE
80 mmHg
SET
FLOW
STOP
HYS
HIHIHI
LO
0
SET
PRESSURE
45 mmHg
SET
FLOW
HYS
HIHIHI
LO
0
ACTUAL PRESSURE
100 150050
INFLOW VOLUME
1.0
INFLOW VOLUME
52
INFLOW VOLUME
1.0
DEFICIT [ml]
0
100
%
0
STOP
Deficit reset
100 ml/min
ACTUAL PRESSURE
100 150050
SET
PRESSURE
80 mmHg
SET
FLOW
STOP
HYS
HIHIHI
LO
0
SET
PRESSURE
45 mmHg
SET
FLOW
HIHIHI
LO
ACTUAL PRESSURE
100 150050
INFLOW VOLUME
1.0
INFLOW VOLUME
52
INFLOW VOLUME
1.0
DEFICIT [ml]
0
100
%
0
0
STOPSTART
Start pump
HYS
100 ml/min
ACTUAL PRESSURE
100 150050
SET
PRESSURE
80 mmHg
SET
FLOW
STOP
HYS
HIHIHI
LO
0
SET
PRESSURE
45 mmHg
SET
FLOW
HYS
HIHIHI
LO
0
ACTUAL PRESSURE
100 150050
INFLOW VOLUME
1.0
INFLOW VOLUME
52
INFLOW VOLUME
1.0
DEFICIT [ml]
52
100
%
0
STOP
100 ml/min
OBJ_DOKU-0000006143-008.fm Seite 71 Mittwoch, 11. November 2020 2:29 14