Safety Notes

16
General Information
en
WARNING!
Air Embolisms
An air embolism can be the result of air contained in the tube set or connected
instrument reaching the patient. Make sure that there is always fluid in the bag
to prevent air from being aspirated into the patient.
3.7 General Laparoscopy Warnings
WARNING!
Connecting the tube
The tube outlet may only be connected to instruments which are intended for in-
tra-abdominal fluid suction and irrigation.
3.8 Arthroscopy Warnings
WARNING!
The height difference between the device and the patient must be adjusted cor-
rectly in order to ensure exact pressure measurement.
WARNING!
Fluid irrigation pumps used in Arthroscopy may cause fluid extravasation into
the surrounding tissue. In severe cases, the resulting pulmonary edema may re-
sult in a serious adverse patient event which may include compartment syn-
drome, neuropraxia, tachycardia, pneumothorax, tissue tearing or nerve
compromise. Should extravasation be observed, it is recommended to reduce
pressure settings and closely observe the excess fluid build up.
WARNING!
The nominal pressure of the described device can be preset to values described
in this manual. Select the optimum pressure based on the patient profile, includ-
ing but not limited to blood pressure, height, weight and age. The manufacturer
recommends pressure settings for different areas of application; however these
are suggestions only and are not intended to substitute or replace the expertise
of the surgeon.
3.9 Hysteroscopy Warnings
WARNING!
When using the scale, follow the operating instructions in this manual.
WARNING!
The deficit and inflow values are lost in case of a power loss or "brownout".
WARNING!
Intrauterine distention is usually possible with pressure values between 35 to
70 mmHg. A pressure above 75 to 80 mmHg is required only in rare cases or if the
patient has an excessively high blood pressure.
OBJ_DOKU-0000006143-008.fm Seite 16 Mittwoch, 11. November 2020 2:29 14