Safety Notes
Using the Hysteroscopy Indication
69
en
Weight > 7 kgIf the weight of the connected scale exceeds the value of 7 kg, the status line de-
picts Scale overloaded! in red letters and a continuous acoustic signal is emitted.
Stop the pump and remove the weight.
Scales defective!If the connected scale is detected as being defective during operation of the de-
vice, the status line depicts the following error message in red: Scale defective! A
short warning beep is emitted 3 times and the symbol is depicted. Safe-
guard the device until checked by an authorized service technician (see Chapter
15.4 "Aesculap Technical Service").
Surgery can be finished but without the scale.
10.7 Preselecting the Nominal Pressure
WARNING!
If the current pressure does not react to an increase of the flow value during sur-
gery, a perforation of the cavum uteri might be the cause. This results in an in-
creased risk for bacteria entering the body. Examine the uterus for injuries.
The nominal pressure can be increased or decreased while the device is being
used or not used. Values may be in the range of 15 to 150 mmHg.
Increasing nominal pressure• Briefly press the [+] symbol to increase the nominal pressure in increments of
5 mmHg. Pressing the [+] symbol longer than 1.5 seconds activates scrolling of
the values in increments of 10 mmHg.
The nominal pressure can also be raised with the help of the remote control
(see Chapter 6.13 "Using the Remote Control (Optional)").
Factory setting: 45 mmHg
When setting a nominal pressure value higher than 100 mmHg, a warning
beep is emitted 1 time and the status line depicts. >100 mmHg?.
ACTUAL PRESSURE
100 150050
SET
PRESSURE
80 mmHg
SET
FLOW
INFLOW VOLUME
1.0
HYS
HIHIHI
LO
SET
PRESSURE
45 mmHg
SET
FLOW
HYS
HIHIHI
LO
ACTUAL PRESSURE
100 150050
Scale overloaded!
STOP
0
INFLOW VOLUME
1.0
DEFICIT [ml]
260
100
%
0
100 ml/min
ACTUAL PRESSURE
100 150050
SET
PRESSURE
80 mmHg
SET
FLOW
STOP
HYS
HIHIHI
LO
0
SET
PRESSURE
45 mmHg
SET
FLOW
HYS
HIHIHI
LO
0
ACTUAL PRESSURE
100 150050
STOP
Scale defective!
INFLOW VOLUME
1.0
INFLOW VOLUME [l]
0.0
100 ml/min
ACTUAL PRESSURE
100 150050
SET
PRESSURE
80 mmHg
SET
FLOW
STOP
HYS
HIHIHI
LO
0
SET
PRESSURE
45 mmHg
SET
FLOW
HIHIHI
LO
ACTUAL PRESSURE
100 150050
INFLOW VOLUME
1.0
INFLOW VOLUME
52
INFLOW VOLUME
1.0
DEFICIT [ml]
0
100
%
0
0
STOPSTART
Start pump
HYS
100 ml/min
OBJ_DOKU-0000006143-008.fm Seite 69 Mittwoch, 11. November 2020 2:29 14