Safety Notes
66
Using the Hysteroscopy Indication
en
10.5 Switching Device on When Using Standard Tube Set
Switching device on without standard tube
set
1. Make sure no tube set is inserted into the tube retainer before switching the
device on.
2. Press the ON switch. The device turns on. The touchscreen displays the fol-
lowing in succession:
Company logo/Device check/Device OK
3. After the successful device check, the available indications are displayed for
selection (see also Chapter 6.3 "Activating a New Indication"). If the device
check is unsuccessful, please read Chapter 6.1.1 "Device Displays after Unsuc-
cessful Completion of Device Check").
4. Press the [HYS] function field.
5. The hysteroscopy indication screen is displayed. The displayed parameters
correspond with the factory settings or the last saved nominal values. The
status line depicts the following: Insert tube set.
Tube set not valid In case an invalid tube set has been inserted, the status line alternately shows the
following: Tube set not valid and Please replace. A short acoustic alarm is emitted
3x.
Device check
LAP
ARTHRO
HYS URO
4.
ACTUAL PRESSURE
100 150050
SET
PRESSURE
80 mmHg
SET
FLOW
STOP
INFLOW VOLUME
1.0
HYS
HIHIHI
LO
0
SET
PRESSURE
45 mmHg
SET
FLOW
HIHIHI
LO
0
ACTUAL PRESSURE
100 150050
Tube set not valid
STOPSTART
100 ml/min
HYS
INFLOW VOLUME [l]
0.0