Safety Notes
Using the Hysteroscopy Indication
65
en
The deficit volume is displayed when the
scale is connected
Once the scale is connected to the device, the differential volume (deficit volume)
is displayed by the device instead of the consumed amount of fluid (volume dis-
play (3)).
10.4 Inserting a Tube Set
Tube types HysteroscopyThe following tube sets can be used for the indication Hysteroscopy (see also
Chapter 20 Article/Accessory List PG130/PG145):
Table 10-1
A standard tube set is not inserted into the tube retainer until the device has
been switched on and the device check has been completed.
CAUTION!
When inserting a tube set already filled with irrigation fluid into the tube retain-
er, it is possible to damage the membranes. Only insert unfilled tube sets to
avoid such damage.
For additional information, see 6.5 "Using the Tube Sets", 6.7 "Inserting a Stan-
dard Tube Set" and 6.9 "Removing a Tube Set".
Fig. 10-2 Hysteroscopy indication screen
displays with connected scale
(1) Indication
(2) Actual pressure display
(3) Differential volume display
(4) Scale connected
(5) Suction function display (only
PG145)
(6) Nominal flow display
(7) [HI] symbol to increase nominal
flow
(8) [Indication selection] symbol
(9) [User menu] symbol
(10) [Save] symbol
(11) [Fluid bag] symbol
(12) [LO] symbol to decrease nominal
flow
(13) Status line for messages
(14) [->0<-] symbol for deficit reset
(15) [+] symbol to increase nominal
pressure
(16) [START]/[STOP] function field
(17) [-] symbol to decrease nominal
pressure
(18) Nominal pressure display
ACTUAL PRESSURE
100 150050
SET
PRESSURE
80
mmHg
SET
FLOW
STOP
HYS
HIHIHI
LO
0
SET
PRESSURE
45 mmHg
SET
FLOW
HIHIHI
LO
0
ACTUAL PRESSURE
100 150050
INFLOW VOLUME
1.0
INFLOW VOLUME
52
INFLOW VOLUME
1.0
DEFICIT [ml]
0
100
%
0
START
Insert tube set
HYS
0
2
100 ml/min
(1) (2) (3) (4)
(5)
(8)(9)(10)(11)(12)(14)
(17)
(18)
(7)
(16)
(15) (13)
(6)
Tube type Tube category
Disposable (one-
time use)
Disposable tube set, Luer lock connector, and tap spikes
Disposable tube set, Luer lock connector, and Care lock
Reusable Reusable tube set, Luer lock connector, and tap spikes
OBJ_DOKU-0000006143-008.fm Seite 65 Mittwoch, 11. November 2020 2:29 14