Safety Notes

Using the Hysteroscopy Indication
63
en
WARNING!
The pressure should be kept as low as possible to allow for a sufficient intrauter-
ine distension and to reduce the forces that could allow fluid, ambient air, and/
or gas into the circulatory system.
WARNING!
Fluid Overload
There is a risk of irrigation fluid reaching the circulatory system of the patient's
soft tissue. This can be affected by distention pressure, flow rate, perforation of
the distended body cavity and duration of the endoscopic surgery. It is critical to
closely monitor the input and outflow of the distending liquid at all times.
WARNING!
Filling the tubing with irrigation fluid and resetting the deficit display to zero are
to be done at the physician’s discretion.
WARNING!
The height difference between the device and the patient must be adjusted cor-
rectly in order to ensure exact pressure measurement.
10.2 Switching Device On
Switch device on1. Make sure no tube set is inserted into the tube retainer before switching the
device on.
2. Press the ON switch. The device turns on. The touchscreen displays the fol-
lowing in succession: Company logo/Device check and Company logo/Device
OK.
3. After the successful device check, the available indications are displayed for
selection (see Chapter 6.3 Activating a New Indication). If the device check is
unsuccessful, please read Chapter 6.1.1 Device Displays after Unsuccessful
Completion of Device Check).
4. Press the [--] function field. The indication screen Hysteroscopy is displayed.
LAP
ARTHRO
HYS URO
4.