9.9 Opening and Configuring the User Menu for the Arthroscopy In-
dication
User menu overviewDevice parameters are displayed and changed in the [ARTHRO settings] menu.
The following chart provides an overview over the possible settings.
1. In the user menu, press the [ARTHRO settings] function field. This opens the
arthroscopy user menu.
2. Press one of the following function fields: [WASH time], [Height differential],
or [Factory settings].
The following is a detailed description of how to change the indication-specific
device parameters.
9.9.1 Setting the WASH Time
Set Wash time1. In the user menu, press the [ARTHRO settings] function field. This opens the
arthroscopy user menu.
2. Press the [WASH time] function field. Select 10, 20, or 30 seconds as the Wash
time.
3. Press the [10], [20], or [30] symbol to increase or decrease the Wash time.
You can now do the following:
4. Press [Save] to save the setting. After saving, the previous menu level is auto-
matically opened.
5. Press [o] to return to the previous menu level without saving.
6. Press [Exit user menu] to return to the screen display of the laparoscopy indi-
cation without saving.
Factory setting: 30 s