Safety Notes

56
Using the Arthroscopy Indication
en
WARNING!
Before every surgery, make sure the remote control does not affect other devices
within the surgical field.
Starting and stopping the WASH function
via infrared remote control
1. The infrared remote control is operational and points towards the front of the
device.
2. Press the [WASH] function field to start the Wash function. Please note that
the Wash function stops automatically after 10, 20, or 30 seconds and the de-
vice returns to the previously specified nominal values. The remaining time of
the WASH function is depicted on the upper right of the screen in the form of
a countdown.
3. Press the [WASH] function field to stop the WASH function before the 10, 20,
or 30 seconds expire. The device automatically returns to the previously spec-
ified nominal values and the screen display after stopping the Wash function.
After stopping the Wash function, it can be restarted again.
PRESSURE
WASH
FLOW
START
STOP
.
OBJ_DOKU-0000006143-008.fm Seite 56 Mittwoch, 11. November 2020 2:29 14