Safety Notes
Using the Arthroscopy Indication
55
en
9.8.1 Using WASH Function via Touchscreen
Follow these steps to activate the WASH function via touchscreen while operat-
ing the device:
Starting and stopping the WASH function
via touchscreen
1. Press the [WASH] function field to start the Wash function. Please note that
the Wash function stops automatically after 10, 20, or 30 seconds and the de-
vice returns to the previously specified nominal values. The remaining time of
the WASH function is depicted on the upper right of the screen in the form of
a countdown.
2. Press the [WASH] function field to stop the WASH function before the 10, 20,
or 30 seconds expire. After stopping the WASH function, it can be restarted.
The device automatically returns to the previously specified nominal values
and the screen display after stopping the Wash function. After stopping the
Wash function, it can be restarted again.
9.8.2 Using WASH Function via Foot Pedal (Optional)
The WASH function can also be activated with a connected foot pedal (see Chap-
ter 20 "Article/Accessory List PG130/PG145").
Starting and stopping the WASH function
via foot pedal
1. The foot pedal must be connected and the device must have recognized the
foot pedal after being switched on.
2. Press the foot pedal to start the Wash function. Please note that the Wash
function stops automatically after 10, 20 or 30 seconds and the device re-
turns to the previously specified nominal values. The remaining time of the
WASH function is depicted on the upper right of the screen in the form of a
countdown. After stopping the Wash function, it can be restarted again.
3. Press the foot pedal to stop the WASH function before the 10, 20, or 30 sec-
onds expire. After stopping the Wash function, it can be restarted again.
9.8.3 Using WASH Function via Remote Control (Optional)
The WASH function can also be activated with the help of the infrared remote
control (see Chapter 20 "Article/Accessory List PG130/PG145").
Place remote control into sterile bag before use (see Chapter 20 "Article/Accesso-
ry List PG130/PG145").
CAUTION!
The remote control can be disinfected by wiping off (do not immerse in liquid).
Do not sterilize the remote control.
ARTHRO
SET
PRESSURE
45 mmHg
SET
FLOW
1.5 l/min
ACTUAL PRESSURE
0 50 100 150 200
STOP WASH
Knee
WASH
30 s
Fig. 9-2 Foot pedal elements
(1) Connection cable
(2) Foot pedal
(2)(1)