Safety Notes

Using the Arthroscopy Indication
51
en
9.5 Selecting Joint Type
Select a jointAfter pressing the [ARTHRO] function field, the following joint selection function
fields are displayed on the touchscreen:
•Knee
Shoulder
Small joint
•Hip
•User 1
•User 2
Enter and save your own user profiles into the function fields [User 1] and [User
2].
1. Press the corresponding function field on the touchscreen to select the joint
type.
2. The display depicts the selected joint type, the set values, and the function
fields used to control the device. The factory default settings are as follows:
Table 9-2
Knee
Small joint
User 1
Shoulder
Hip
User 2
ARTHRO
Joint type Nominal pressure
factory setting
Nominal flow fac-
tory setting
Small joints 35 mmHg 0.7 l/min
Knee joint 45 mmHg 1.0 l/min
Shoulder joint 50 mmHg 1.5 l/min
Hip joint 60 mmHg 1.0 l/min
User 1 50 mmHg 1.5 l/min
User 2 50 mmHg 1.5 l/min
OBJ_DOKU-0000006143-008.fm Seite 51 Mittwoch, 11. November 2020 2:29 14