Safety Notes
50
Using the Arthroscopy Indication
en
9.4 Switching Device on When Using Day Patient Set
Switch device on with inserted day tube a. If the device is switched on and a new day set is inserted, the status field of
the display depicts the following: Remove tube set. The day tube must be re-
moved.
b. If the device is switched on and a day set used during the previous surgery is
inserted, the status field of the display depicts the following consecutive
messages: Company logo/Device check and Company logo/Device OK. The
day tube may remain in place.
c. If the day set has become contaminated during the previous surgery, the day
set must be replaced. The start screen opens and depicts the following: Day
tube contaminated, Please replace.
1. After the successful device check, the available indications are displayed for
selection (see also Chapter 6.3 "Activating a New Indication"). If the device
check is unsuccessful, please follow the instructions in Chapter 6.1.1 "Device
Displays after Unsuccessful Completion of Device Check").
2. Press the [ARTHRO] function field. The joint selection is displayed (see Chap-
ter 9.5 "Selecting Joint Type").
3. Press the desired function field, e.g., [Knee].
4. The arthroscopy indication screen is displayed. The displayed parameters cor-
respond with the factory settings or the last saved nominal values. A newly
inserted day tube is validated (see 9.3 "Switching Device on When Using Stan-
dard Tube Set", item 6). The status line depicts the following: Day tube insert-
ed.
Tube set used up In case a used up tube set has been inserted, a warning beep is emitted 3 times
and the status line alternately shows the following: Tube set used up and Please
replace.
Remove tube set
Day tube contaminated
Please replace
ARTHRO
Knee
Small joint
User 1
Shoulder
Hip
User 2
ARTHRO
ARTHRO
SET
PRESSURE
45 mmHg
SET
FLOW
1.5 l/min
ACTUAL PRESSURE
0 50 100 150 200
START
Day tube inserted
Knee
ARTHRO
SET
PRESSURE
45 mmHg
SET
FLOW
1.5 l/min
ACTUAL PRESSURE
0 50 100 150 200
START
Tube set used up
Knee
OBJ_DOKU-0000006143-008.fm Seite 50 Mittwoch, 11. November 2020 2:29 14