Safety Notes
Using the Arthroscopy Indication
49
en
For additional information see 6.5 "Using the Tube Sets", 6.7 "Inserting a Stan-
dard Tube Set", 6.8 "Inserting the Day Patient Tube Set" and 6.9 "Removing a Tube
Set".
9.3 Switching Device on When Using Standard Tube Set
Switching device on without standard tube
set
1. Make sure no tube set is inserted into the tube retainer before switching the
device on.
2. Press the ON switch. The device turns on. The touchscreen displays the fol-
lowing in succession: Company logo/Device check and Company logo/Device
OK.
3. After the successful device check, the available indications are displayed for
selection (see also Chapter 6.3 "Activating a New Indication"). If the device
check is unsuccessful, please read Chapter 6.1.1 "Device Displays after Unsuc-
cessful Completion of Device Check").
4. Press the [ARTHRO] function field. The joint selection is displayed (see Chap-
ter 9.5 "Selecting Joint Type").
5. Press the desired function field, e.g., [Knee].
6. The arthroscopy indication screen is displayed. The displayed parameters cor-
respond with the factory settings or the last saved nominal values. The status
line depicts the following: Insert tube set.
Tube set not validIn case an invalid tube set has been inserted, the status line alternately shows the
following: Tube set not valid and Please replace. A short acoustic alarm is emitted
3x.
Device check
LAP
ARTHRO
HYS URO
4.
Knee
Small joint
User 1
Shoulder
Hip
User 2
ARTHRO
ARTHRO
SET
PRESSURE
45 mmHg
SET
FLOW
1.5 l/min
ACTUAL PRESSURE
0 50 100 150 200
START
Insert tube set
Knee
ARTHRO
SET
PRESSURE
45 mmHg
SET
FLOW
1.5 l/min
ACTUAL PRESSURE
0 50 100 150 200
START
Tube set not valid
Knee
OBJ_DOKU-0000006143-008.fm Seite 49 Mittwoch, 11. November 2020 2:29 14