9.1 Device-inherent Dangers: Arthroscopy Indication
WARNING!
The height difference between the device and the patient must be adjusted cor-
rectly in order to ensure exact pressure measurement.
WARNING!
Fluid irrigation pumps used in Arthroscopy may cause fluid extravasation into
the surrounding tissue. In severe cases, the resulting pulmonary edema may re-
sult in a serious adverse patient event which may include compartment syn-
drome, neuropraxia, tachycardia, pneumothorax, tissue tearing or nerve
compromise. Should extravasation be observed, it is recommended to reduce
pressure settings and closely observe the excess fluid build up.
WARNING!
The nominal pressure of the described device can be preset to values described
in this manual. Select the optimum pressure based on the patient profile, includ-
ing but not limited to blood pressure, height, weight and age. The manufacturer
recommends pressure settings for different areas of application; however these
are suggestions only and are not intended to substitute or replace the expertise
of the surgeon.
9.2 Inserting a Tube Set
Tube types Arthroscopy One of the following two tube sets can be specified in the indication Arthroscopy:
Standard tube set (disposable and reusable)
Day patient set (in short: Day set)
A standard tube set is not inserted into the tube retainer until the device has
been switched on and the device check has been completed. The day set, howev-
er, remains in the tube retainer after surgery (unless it must be replaced).
The following tube sets can be used for the indication Arthroscopy (see also
Chapter 20 Article/Accessory List PG130/PG145):
Table 9-1
CAUTION!
When inserting a tube set already filled with irrigation fluid into the tube retain-
er, it is possible to damage the membranes. Only insert unfilled tube sets to
avoid such damage.
Tube type Tube category
Disposable (one-
time use)
Disposable tube set, Luer lock connector, and tap spikes
Disposable tube set, Luer lock connector, and Care lock
Reusable Reusable tube set, Luer lock connector, and tap spikes
Day patient set Day patient set with Luer lock connector and tap spikes,
consisting of day tube and patient tube
Day patient set with Luer lock connector and Care lock,
consisting of day tube and patient tube
Day tube Day tube with Luer lock connector and tap spikes
Day tube with Luer lock connector and Care lock
OBJ_DOKU-0000006143-008.fm Seite 48 Mittwoch, 11. November 2020 2:29 14