Safety Notes
46
Using the Laparoscopy Indication
en
8.7.1 Setting the Flow Level Standard/Flow Level High
Set the flow level Standard/flow level High
1. In the user menu, press the [LAP settings] function field. This opens the lapa-
roscopy user menu.
2. Press the [Flow level Standard]/[Flow level High] function field. You can set
values between 1.0 and 2.5 l/min for the Standard flow level and a value be-
tween 2.5 and 3.5 l/min for the High flow level.
3. Press the [-] or [+] symbol to increase or decrease the nominal flow.
You can now do the following:
4. Press [Save] to save the setting. After saving, the previous menu level is auto-
matically opened.
5. Press [o] to return to the previous menu level without saving.
6. Press [Exit user menu] to return to the screen display of the laparoscopy indi-
cation without saving.
Factory setting flow level Standard: 1.5 l/min
Factory setting flow level High: 2.5 l/min
8.7.2 Resetting to Factory Settings
Reset to factory settings 1. In the user menu, press the [LAP settings] function field. This opens the lapa-
roscopy user menu.
2. Press the [Factory settings] function field to reset the LAP device parameters
to the factory settings.
You can now do the following:
3. Press [Save] to reset to the factory settings. After saving, the previous menu
level is automatically opened.
4. Press [o] to return to the previous menu level without saving.
5. Press [Exit user menu] to return to the screen display of the laparoscopy indi-
cation without saving.
Menu
Flow level Standard
1,5 l/min
3.
6.
4.
5.
Menu
Factory settings
Press Save button to reset all
indication presets to factory defaults
according to manual
5.
3.
4.